Related compounds
Mobile phase and Chromatographic system
Proceed as directed in the Assay.
Test solution
Transfer about 38 mg of Betahistine Hydrochloride, accurately weighed, to a 100-mL volumetric flask, dissolve in and dilute with Mobile phase to volume, and mix.
Procedure
Inject about 10 µL of the
Test solution into the chromatograph, record the chromatogram, and measure the peak responses. Calculate the percentage of each impurity in the portion of Betahistine Hydrochloride taken by the formula:
100F(ri /rs),
in which
F is the response factor of the respective impurity (see
Table 1) and 1.0 for all other peaks;
ri is the peak response for each impurity; and
rs is the sum of the responses of all of the peaks, adjusted for the relative response factor.
Table 1
Impurity Name |
Relative Retention Time |
Response Factor (F) |
Limit (%) |
2-(2-Hydroxyethyl)pyridine |
0.3 |
0.5 |
0.2 |
2-Vinylpyridine |
0.4 |
0.4 |
0.2 |
N-Methyl-N,N-bis(2-
pyridin-2-yl-ethyl)amine |
2.4 |
1.4 |
0.2 |
In addition to not exceeding the limits for impurities in
Table 1, not more than 0.1% of any other individual impurity is found; and not more than 0.5 % of total impurities is found.
Assay
Ammonium acetate buffer
Dissolve about 0.69 g of ammonium acetate in 1000 mL of water. Adjust with glacial acetic acid to a pH of 4.7.
Mobile phase
Prepare a filtered and degassed mixture of 350 mL of acetonitrile and 650 mL of
Ammonium acetate buffer, containing 2.88 g of sodium lauryl sulfate. Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Standard preparation
Dissolve an accurately weighed quantity of
USP Betahistine Hydrochloride RS in
Mobile phase to obtain a solution having a known concentration of about 0.38 mg per mL.
Assay preparation
Transfer about 38 mg of Betahistine Hydrochloride, accurately weighed, to a 100-mL volumetric flask, dissolve in and dilute with Mobile phase to volume, and mix.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 254-nm detector and 3.0-mm × 15-cm column that contains packing L1. The column temperature is maintained at 40
. The flow rate is about 0.5 mL per minute. Chromatograph the
Standard preparation, and record the peak responses as directed for
Procedure: the column efficiency is not less than 5000 theoretical plates; the tailing factor for the betahistine peak is not more than 2.0; and the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
Separately inject equal volumes (about 10 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in mg, of C
8H
12N
2·2HCl in the portion of Betahistine Hydrochloride taken by the formula:
100C(rU /rS),
in which
C is the concentration, in mg per mL, of
USP Betahistine Hydrochloride RS in the
Standard preparation; and
rU and
rS are the peak responses obtained from the
Assay preparation and the
Standard preparation, respectively.