Packaging, storage, and labeling
Requirements for packaging, storage, and labeling do not apply; Rubidium Chloride Rb 82 Injection is obtained by elution from the generator and is administered by direct infusion.
Bacterial endotoxins 85
It contains not more than 175/
V USP Endotoxin Unit per mL of the Injection, when compared with the USP Endotoxin RS, in which
V is the maximum recommended total dose, in mL, at the expiration date or time.
Radionuclide identification (see Radioactivity 821)
[NOTEPerform this test quickly, because of the rapid decay of the
82Rb.
] The gamma-ray spectrum of eluted
82Rb exhibits photopeaks at 511 and 777 keV.
Radionuclidic purity
Using a suitable counting assembly (see
Selection of a Counting Assembly under
Radioactivity 821), determine the radioactivity of each radionuclidic impurity, in kBq per MBq (or µCi per mCi), of Rb 82 in the generator eluate by use of a calibrated system as directed under
Radioactivity 821.
[NOTEFor the following tests, use the generator eluate containing
82Rb that has been allowed to decay for 1 hour after the end of elution.
]
Sr 82 and Rb 83
Obtain a gamma-ray spectrum of the hour-old eluate, and measure the activities of the radionuclidic impurities directly from the spectrum. Sr 82 exhibits photopeaks at 511 and 777 keV and decays with a radioactive half-life of 25.5 days. Rb 83 exhibits a photopeak at 530 keV and decays with a radioactive half-life of 86.2 days. The activity levels of Sr 82 and Rb 83 are not more than 0.02 kBq per MBq (0.02 µCi per mCi) and not more than 0.05 kBq per MBq (0.05 µCi per mCi) of Rb 82 at the end of elution, respectively.
Sr 85
Obtain a gamma spectrum of the hour-old eluate, and, using the same system and geometry, obtain a gamma spectrum of a pure Rb 82 specimen (generator eluate containing 82Rb taken within 10 minutes of elution). Sr 85 exhibits a major photopeak at 514 keV and decays with a radioactive half-life of 64.8 days. Sr 85 may be determined by subtraction of the 511 and 777 keV peaks in the pure Rb 82, from the 511514 keV and 777 keV peaks in the hour-old eluate. The activity level of Sr 85 is not more than 0.2 kBq per MBq (0.2 µCi per mCi) of Rb 82 at the end of elution.
Other gamma-ray emitters
The total of other gamma-ray emitting radionuclidic impurities does not exceed 0.005 kBq per MBq (0.005 µCi per mCi) of Rb 82 at the end of elution.
Chemical purity
Electrolyte solution
Transfer 107 g of ammonium chloride, 25 g of gelatin, and 42 mL of hydrochloric acid to a 500-mL volumetric flask. Add about 450 mL of water, and sonicate until a clear solution is obtained. Dilute with water to volume, and mix.
Tin stock standard solution
Dissolve 100 mg of metallic tin (Sn), accurately weighed, in 10 mL of dilute hydrochloric acid (1 in 2), and dilute with water to 100 mL.
Tin standard solution A
Transfer 0.5 mL of Tin stock standard solution to a 50-mL volumetric flask and dilute with 0.1 N hydrochloric acid to volume.
Tin standard solution B
Transfer 1.0 mL of Tin standard solution A to a 50-mL volumetric flask. Add 10.0 mL of 0.9% sodium chloride solution, dilute with Electrolyte solution to volume, and mix.
Test solution
Obtain a 50-mL eluate from the generator, and allow to stand for at least 1 hour to allow for the complete decay of 82Rb. Transfer 10.0 mL of the eluate to a 50-mL volumetric flask, dilute with the Electrolyte solution to volume, and mix.
Procedure
Transfer a portion of the
Test solution to a polarographic cell, and deaerate by bubbling nitrogen through the solution for 5 minutes. Insert the dropping mercury electrode of a suitable polarograph, and obtain the differential pulse polarogram from
0.15 to
0.75 volts, at a current range of 0.5 µA, using a saturated calomel electrode as the reference electrode and a platinum wire as the auxiliary electrode (see
Polarography 801). Similarly, transfer a portion of the
Tin standard solution to a polarographic cell and obtain the polarogram. A peak at
0.52 volts indicates the presence of tin. The peak height of the
Test solution is not greater than that of the
Tin standard solution (1 µg per mL).
Other requirements
It meets the requirements under
Injections 1, except that the Injection may be distributed or dispensed prior to completion of the test for
Sterility, the latter test being started on the day of final manufacture, and except that it is not subject to the recommendation for
Volume in Container under
Injections 1.