Assay
Mobile phase
Mix 300 mL of a 1 in 50 solution of ammonium hydroxide in isopropyl alcohol and 700 mL of n-hexane. Filter through a 0.5-µm filter before using.
Standard preparation
Using an accurately weighed quantity of
USP Pilocarpine Hydrochloride RS, prepare a solution having a known concentration of about 1.6 mg per mL.
Assay preparation
Transfer an accurately measured volume of Ophthalmic Solution, equivalent to about 80 mg of pilocarpine hydrochloride, to a 50-mL volumetric flask. Dilute with methanol to volume, and mix.
Chromatographic system
(see
Chromatography 621)The liquid chromatograph is equipped with a 220-nm detector and a 4.6-mm × 25-cm column that contains packing L3. The flow rate is about 2 mL per minute. Chromatograph three replicate injections of the
Standard preparation, and record the peak responses as directed for
Procedure: the relative standard deviation is not more than 2.0%.
Procedure
Separately inject equal volumes (about 10 µL) of the
Standard preparation and the
Assay preparation into the chromatograph by means of a suitable microsyringe or sampling valve, record the chromatograms, and measure the responses for the major peaks. The retention time is about 16 minutes for pilocarpine hydrochloride. Calculate the quantity, in mg, of C
11H
16N
2O
2·HCl in each mL of the Ophthalmic Solution taken by the formula:
50(C / V)(rU / rS),
in which
C is the concentration, in mg per mL, of
USP Pilocarpine Hydrochloride RS in the
Standard preparation,
V is the volume, in mL, of Ophthalmic Solution taken, and
rU and
rS are the peak responses obtained from the
Assay preparation and the
Standard preparation, respectively.