Chromatographic purity
Solution A
Prepare a filtered and degassed solution of 1 g of monobasic potassium phosphate in 1000 mL of water, adjust with 45% potassium hydroxide to a pH of 5.5, and mix. Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Solution B
Use filtered and degassed acetonitrile.
Solvent
Prepare a mixture of acetonitrile and Solution A (80:20).
Standard solution
Dissolve an accurately weighed quantity of
USP Hydrocortisone Butyrate RS in
Solvent to obtain a solution having a known concentration of about 0.5 mg per mL.
Test solution
Transfer about 50 mg of Hydrocortisone Butyrate, accurately weighed, to a 50-mL volumetric flask, dissolve in and dilute with Solvent to volume, and mix.
Chromatographic system
(see
Chromatography 621)The liquid chromatograph is equipped with a 254-nm detector and a 4.6-mm × 10-cm column that contains 3-µm packing L1. The flow rate is about 2.0 mL per minute. The chromatograph is programmed as follows.
Time (minutes) |
Solution A
(%) |
Solution B
(%) |
Elution |
0 |
80 |
20 |
equilibration |
0-12.5 |
80®35 |
20®65 |
linear gradient |
12.5-15.5 |
35 |
65 |
isocratic |
15.5-20.5 |
35®80 |
65®20 |
linear gradient |
20.5-22.5 |
80 |
20 |
re-equilibration |
Chromatograph the
Standard solution and the
Test solution, and record the peak responses as directed for
Procedure: the resolution,
R, between hydrocortisone butyrate and any impurity is not less than 1.0; and the column efficiency is not less than 10,000 theoretical plates.
Procedure
Separately inject equal volumes (about 5 µL) of the
Test solution and the
Standard solution into the chromatograph, record the chromatograms, and measure the areas for the major peaks, disregarding any peak having a percentage of 0.05% or less. Calculate the percentage of each impurity in the portion of Hydrocortisone Butyrate taken by the formula:
100(CS / CU)(ri / rS),
in which
CS is the concentration, in mg per mL, of
USP Hydrocortisone Butyrate RS in the
Standard solution; CU is the concentration, in mg per mL, of Hydrocortisone Butyrate in the
Test solution; ri is the peak area for each impurity obtained from the
Test solution; and
rS is the peak area for hydrocortisone butyrate obtained from the
Standard solution: not more than 1.0% of any individual impurity is found; and not more than 2.0% of total impurities is found.
Assay
Mobile phase
Prepare a filtered and degassed mixture of water, acetonitrile, and glacial acetic acid (124:76:1). Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Solvent
Prepare a mixture of tetrahydrofuran and glacial acetic acid (1000:1).
Diluting solution
Prepare a filtered solution of methanol, water, and glacial acetic acid (500:500:1).
Standard preparation
Dissolve an accurately weighed quantity of
USP Hydrocortisone Butyrate RS in
Solvent, and dilute quantitatively, and stepwise if necessary, with
Solvent to obtain a solution having a known concentration of about 0.1 mg per mL. Transfer 10.0 mL of this solution to a 50-mL volumetric flask, dilute with
Diluting solution to volume, and mix.
System suitability solution
Dissolve suitable quantities of propyl 4-hydroxybenzoate and
USP Hydrocortisone Butyrate RS in
Solvent, and dilute quantitatively, and stepwise if necessary, with
Solvent to obtain a solution having known concentrations of about 0.1 mg of each per mL. Transfer 10 mL of this solution to a 50-mL volumetric flask, dilute with
Diluting solution to volume, and mix.
Assay preparation
Transfer about 50 mg of Hydrocortisone Butyrate, accurately weighed, to a 50-mL volumetric flask, dissolve in and dilute with Solvent to volume, and mix. Transfer 5.0 mL of this solution to a 50-mL volumetric flask, dilute with Solvent to volume, and mix. Transfer 10.0 mL of this solution to a 50-mL volumetric flask, dilute with Diluting solution to volume, and mix.
Chromatographic system
(see
Chromatography 621)The liquid chromatograph is equipped with a 254-nm detector and a 3.0-mm × 10-cm column that contains packing L1. The flow rate is about 1 mL per minute. Chromatograph the
System suitability solution, and record the peak responses as directed for
Procedure: the relative retention times are about 0.7 for propyl 4-hydroxybenzoate and 1.0 for hydrocortisone butyrate; and the resolution,
R, between propyl 4-hydroxybenzoate and hydrocortisone butyrate is not less than 4.0. Chromatograph the
Standard preparation, and record the peak responses as directed for
Procedure: the column efficiency is not less than 4000 theoretical plates; the tailing factor is not more than 1.6; and the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
Separately inject equal volumes (about 5 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in mg, of C
25H
36O
6 in the portion of Hydrocortisone Butyrate taken by the formula:
2500C(rU / rS),
in which
C is the concentration, in mg per mL, of
USP Hydrocortisone Butyrate RS in the
Standard preparation; and
rU and
rS are the peak responses obtained from the
Assay preparation and the
Standard preparation, respectively.