Related compounds
Spray reagent
[CautionAvoid contact with o-tolidine. Prepare and use this Spray reagent in a well-ventilated hood.] Dissolve 50 mg of o-tolidine in 100 mL of alcohol, and mix.
Chlorine chamber
Transfer 1.5 g of potassium permanganate to a 100-mL beaker, dissolve in and dilute with water to volume, and mix. Transfer 25 mL of this solution to a beaker, and place the beaker inside a chromatographic chamber. Pipet 10 mL of hydrochloric acid into the beaker, and cover the chamber.
Developing solvent system
Prepare a fresh mixture of ethyl acetate, glacial acetic acid, and acetonitrile (70:25:3).
Standard solutions
Dissolve accurately weighed quantities of
USP Guanfacine Hydrochloride RS and guanidine hydrochloride in methanol to obtain a solution having a known concentration of 0.4 mg each of
USP Guanfacine Hydrochloride RS and guanidine hydrochloride per mL. Quantitatively dilute this solution with methanol to obtain
Standard solutions having the following compositions:
Standard Solution |
Dilution |
Concentration (µg RS and Guanidine Hydrochloride per mL) |
Percentage (% for Comparison with Test Specimen) |
1 |
(undiluted) |
400 |
2.0 |
2 |
(1 in 2) |
200 |
1.0 |
3 |
(1 in 4) |
100 |
0.5 |
4 |
(1 in 8) |
50 |
0.25 |
Test solution
Dissolve an accurately weighed quantity of Guanfacine Hydrochloride in methanol to obtain a solution having a concentration of about 20 mg per mL.
Procedure
Use a thin-layer chromatographic plate (see
Chromatography 621 coated with a 0.25-mm layer of chromatographic silica gel. Prewash the plates by placing in a chromatographic chamber saturated with
Developing solvent system. Remove the plates from the chamber, and allow to dry. Separately apply 10 µL each of the
Standard solutions and the
Test solution to the chromatographic plate. Allow the spots to dry, and develop the chromatogram in
Developing solvent until the solvent front has moved about three-fourths of the length of the plate. Remove the plate from the chamber, mark the solvent front, and allow the plate to air-dry for about 1 hour. Examine the plate under short-wavelength UV light. Place the dried plate in the
Chlorine chamber for 15 minutes, remove, and allow the excess chlorine to evaporate by air drying for 5 minutes. Spray the plate with
Spray reagent, and examine: any spot due to guanidine hydrochloride observed in the chromatogram of the
Test solution is not greater in size or intensity than the guanidine hydrochloride spot obtained from
Standard solution 3 (0.5%); no other individual impurity spot observed in the chromatogram of the
Test solution is greater in size or intensity than the guanfacine hydrochloride spot obtained from
Standard solution 4 (0.25%); and the sum of all impurities found, including guanidine hydrochloride, is not more than 1.0%.
Chromatographic purity
Spray reagent 1
Prepare a mixture of tertiary butyl alcohol and water (9:1).
Spray reagent 2
Dissolve 5 g of 4,4¢-tetramethyldiaminodiphenylmethane in 20 mL of glacial acetic acid, add 10 mL of water, and mix (Solution 1). Dissolve 6 g of potassium iodide in 120 mL of water, and mix (Solution 2). Dissolve 0.3 g of ninhydrin in 10 mL of glacial acetic acid, dilute with water to 100 mL, and mix (Solution 3). Mix Solution 1 and Solution 2, and add 9 mL of Solution 3.
Developing solvent system
Prepare a fresh mixture of hexanes, diisopropyl ether, toluene, and glacial acetic acid (60:30:5:3).
Reference solutions
Dissolve an accurately weighed quantity of 2,6-dichlorophenylacetic acid in a mixture of methanol and water (9:1) to obtain a solution having a concentration of 1 mg per mL (Reference solution 1). Quantitatively dilute this solution with a mixture of methanol and water (9:1) to obtain Reference solution 2 and Reference solution 3 having known concentrations of 0.5 and 0.25 mg per mL of 2,6-dichlorophenylacetic acid, respectively.
Test solution
Prepare a solution of Guanfacine Hydrochloride in a mixture of methanol and water (9:1), containing 100 mg per mL.
Procedure
Use a thin-layer chromatographic plate (see
Chromatography 621) coated with a 0.25-mm layer of chromatographic silica gel. Prewash the plates by placing in a chromatographic chamber saturated with
Developing solvent system. Remove the plates from the chamber, and allow to dry. Separately apply 25 µL of each of the
Reference solutions and the
Test solution to the chromatographic plate. Allow the spots to dry, and develop the chromatograms in the
Developing solvent system until the solvent front has moved about three-fourths of the length of the plate. Remove the plate from the chamber, mark the solvent front, and allow the plate to air-dry for 30 minutes. Examine the plate under short-wavelength UV light. Spray the plate with
Spray reagent 1, wait for 1 minute, and then spray with
Spray reagent 2. Place the wet plate under short-wavelength UV light for 10 minutes, remove, and observe under white light: no spot observed in the chromatogram of the
Test solution, other than that due to guanfacine hydrochloride, is greater in size or intensity than the principal spot obtained from
Reference solution 2 (0.5%); and the sum of all impurities found is not more than 1.0%.
Assay
Dilute phosphoric acid
Prepare a mixture of water and phosphoric acid (4:1).
Buffer solution
Dissolve 68 g of monobasic potassium phosphate in water, dilute with water to 1000 mL, and mix. Dilute 100 mL of this solution with water to 1000 mL, add 5 mL of triethylamine, mix, and adjust with Dilute phosphoric acid to a pH of 3.0.
Mobile phase
Prepare a filtered and degassed mixture of
Buffer solution and acetonitrile (79:21). Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Standard preparation
Dissolve an accurately weighed quantity of
USP Guanfacine Hydrochloride RS in a mixture of acetonitrile and water (3:1) to obtain a solution having a known concentration of about 1 mg of
USP Guanfacine Hydrochloride RS per mL. Transfer 2.0 mL of this solution to a 50-mL volumetric flask, dilute with
Mobile phase to volume, and mix.
Assay preparation
Transfer an accurately weighed quantity of about 50 mg of Guanfacine Hydrochloride to a 50-mL volumetric flask, dissolve in and dilute with a mixture of acetonitrile and water (3:1) to volume, and mix. Transfer 2.0 mL of this solution to a 50-mL volumetric flask, dilute with Mobile phase to volume, and mix.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 220-nm detector and a 4.6-mm × 15-cm column that contains packing L1. The flow rate is about 1 mL per minute. Chromatograph the
Standard preparation, and record the responses as directed for
Procedure: the capacity factor,
k¢, is between 2 and 5; the column efficiency is not less than 1500 theoretical plates; the tailing factor is not more than 2; and the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
Separately inject equal volumes (about 20 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the areas for the major peaks. Calculate the quantity, in mg, of C
9H
9Cl
2N
3O·HCl in the portion taken by the formula:
1.25C(rU/rS),
in which
C is the concentration, in µg per mL, of
USP Guanfacine Hydrochloride RS in the
Standard preparation; and
rU and
rS are the guanfacine hydrochloride peaks obtained from the
Assay preparation and the
Standard preparation, respectively.