Chromatographic system
(see
Chromatography 621)The gas chromatograph is equipped with a thermal conductivity detector and a 1.8-m × 4-mm glass column (pretreated with 10% dimethyldichlorosilane in toluene) packed with 10% phase G16 on 60- to 80-mesh support S1AB. The column is maintained at about 160
, and the injector and the detector are maintained at about 200
. The carrier gas is dry helium, flowing at a rate of about 30 mL per minute. Chromatograph the
Resolution solution, and record the peak responses as directed for
Procedure: the relative retention times for toluene,
-chlorovinylethyl ketone (if present),
Z-ethchlorvynol, and
E-ethchlorvynol are about 0.1, 0.2, 0.8, and 1.0, respectively; the resolution,
R, between the
Z- and
E-ethchlorvynol peaks is not less than 1.0; and the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
[NOTEUse peak areas where peak responses are indicated.
] Inject about 3 µL of Ethchlorvynol into the chromatograph, record the chromatogram, and measure the responses for all of the peaks. Calculate the percentage of
E-ethchlorvynol (C
7H
9ClO) in the specimen of Ethchlorvynol taken by the formula:
100re / rt,
in which
re is the response of the
E-ethchlorvynol peak obtained in the chromatogram of the Ethchlorvynol and
rt is the sum of the responses of all of the peaks observed in the chromatogram.