Microsphere size and concentration
Electrolyte solution
Use filtered and buffered saline electrolyte solution.
1
Diluent
Prepare a solution that contains, in each L of water, 1.5 g of sodium lauryl sulfate and 0.1 g of thimerosal. Prior to use, pass the solution through a 0.2-µm nylon filter. [NOTEThe Diluent is to be used exclusively to prepare the Reference stock solution described below; it must not be used to prepare the Test solution.]
Reference stock solution
Transfer a quantity of NIST traceable microspheres suspension containing about 0.5 g of microspheres directly into a tared centrifuge tube equipped with a cap, and weigh.
2 Using the density and concentration of the microspheres obtained from the Certificate of Analysis, calculate the volume occupied by the microspheres and the number of microspheres in the portion taken. Calculate the target total volume, in mL, by dividing the number of microspheres in the portion taken by the target concentration of 2.0 × 10
8 microspheres per mL. Calculate the target
Diluent volume by subtracting the volume occupied by the microspheres from the target total volume. Transfer the target volume of
Diluent to the centrifuge tube containing the portion of microspheres taken, and mix the tube vigorously for 1 hour. The prepared
Reference stock solution, which contains a suspension of microspheres with a target mean particle diameter of 5.2 µm and a concentration of 2.0 × 10
8 microspheres per mL, is divided into smaller containers and stored at 5
.
Reference solution
Equilibrate the Reference stock solution to room temperature, and mix thoroughly. Immediately transfer 20 µL of the Reference stock solution to a beaker containing 200 mL of Electrolyte solution, mix, and analyze immediately.
Blank solution
Use 200 mL of Electrolyte solution.
Test solution
Allow the Injectable Suspension to equilibrate to room temperature. Invert the vial, and gently rotate to resuspend the microspheres. [NOTEAfter resuspension, the contents should appear as a homogeneous, opaque, milky-white suspension.] Immediately withdraw a 20-µL aliquot, transfer to a beaker containing 200 mL of Electrolyte solution, mix, and analyze immediately.
Test apparatus
Use a multichannel particle analyzer that operates on the electrical zone-sensing principle.
3 The analyzer is fitted and calibrated with an aperture tube having a 50-µm orifice. The multichannel particle analyzer is equipped with software capable of data-smoothing, data extrapolation, distribution graphing, and data conversion. Analyze the
Blank solution, the
Reference solution, and the
Test solution as directed for
Procedure: the total count in the
Blank solution is not more than 500; the mean particle diameter of microspheres in the
Reference solution is within 5% of the mean particle diameter of microspheres in the
Reference stock solution; the concentration of the
Reference solution is within 10% of the concentration of the
Reference stock solution; and the coincidence effect in the analysis of the
Test solution is not more than 5%.
Procedure
Rinse the orifice of the aperture tube with
Electrolyte solution before and after analyzing each preparation. Place the
Blank solution in the apparatus, and adjust the vacuum on the sample stand so that the counting begins about 12 seconds after the analyzer is set to the counting position. Set the data acquisition to stop when one of the following conditions is met: preset length of time, preset volume, preset number of counts in any channel, or total counts. Collect the count versus the channel data for the
Blank solution, and analyze using the data-smoothing, data extrapolation, distribution graphing, and data conversion features of the system software. In the same manner, analyze the
Reference solution and the
Test solution. The
Test solution data are normalized and expressed as the number of microspheres per mL: the concentration of microspheres is between 5.0 × 10
8 and 8.0 × 10
8 per mL. Calculate the percentage of microspheres less than 10 µm in size in the portion of Injectable Suspension taken by the formula:
100(PA / PB),
in which
PA is the number of microspheres in the 1- to 10-µm size range, and
PB is the number of microsphere particles in the 1- to 32-µm size range. Not fewer than 93% of microsphere particles is smaller than 10 µm.
Container headspace content
Reference solutions
Use 99% perflutren reference standard
4 (electronic grade perflutren gas of at least 99 molar % purity) and 60% perflutren reference standard
4 (an electronic grade gas in air mixture containing 60 molar % perflutren gas).
Blank solution
Use ambient air.
Test solution
Use gas from the container headspace.
Chromatographic system (see Chromatography 621)
The gas chromatograph is equipped with a thermal conductivity detector and a 0.53-mm × 25-m fused-silica (porous layer open tubular) column coated with Al
2O
3 / KCl (aluminum oxide deactivated with potassium chloride).
5 The carrier gas is helium with a flow rate adjusted to obtain a retention time of about 1.5 to 1.8 minutes for perflutren. The column temperature is maintained at about 65
, the injection port temperature is maintained at about 130
, and the detector temperature is maintained at about 180
. Chromatograph the
Reference solutions as directed for
Procedure: the resolution,
R, between perflutren and air is not less than 2; and the relative standard deviation for replicate injections is not more than 5%. The measured value for the 99% perflutren reference standard is within 5% of the nominal value.
Procedure
Using a gas-tight syringe, separately inject 10 µL of the Blank solution, the Reference solutions, and the Test solution into the chromatograph. Record the chromatograms, and measure the responses for the major peaks. The percentages of perflutren in the 99% perflutren reference standard and in the Test solution are calculated by comparing the peak areas in each with the peak areas obtained from the 60% perflutren reference standard. The container headspace of Injectable Suspension contains not less than 60% of perflutren gas.
Microsphere perflutren content
Reference stock solutions
The 97% decafluorobutane reference standard is decafluorobutane gas of at least 97 molar % purity.
6 The 5% decafluorobutane5% perflutren reference standard is a mixture containing 5 molar % decafluorobutane gas and 5 molar % perflutren gas in air.
7 The 99% perflutren reference standard is electronic grade perflutren gas of at least 99 molar % purity.
4
Analysis vial
Transfer 100 µL of 97% decafluorobutane reference standard gas and 100 µL of glacial acetic acid to a 2-mL vial equipped with a septum cap.
Reference solution
Transfer 100 µL of 99% perflutren reference standard and 0.75 mL of 1% Albumin Human to an Analysis vial, and incubate by mixing for at least 3 hours.
Test solution
Allow a vial of the Injectable Suspension to equilibrate to room temperature. Invert the vial, and gently rotate to resuspend the microspheres. [NOTEAfter resuspension, the contents of the vial should appear as a homogeneous, opaque, milky-white suspension.] Withdraw 0.75 mL of Injectable Suspension, and transfer to another Analysis vial. Incubate the Test solution by mixing for at least 3 hours.
Chromatographic system (see Chromatography 621)
Prepare as directed for
Container headspace content. The carrier gas is helium with a flow rate adjusted to obtain the following retention times: 1.0 to 1.1 minutes for air, 1.3 to 1.5 minutes for perflutren, and 1.5 to 2.5 minutes for decafluorobutane. The column temperature is maintained at about 85
, and then after elution of the perflutren the temperature is increased at a rate of 50
per minute to 120
, and maintained at 120
for 2 minutes. The injection port temperature is maintained at about 130
, and the detector temperature is maintained at about 180
. Chromatograph the 5% decafluorobutane5% perflutren reference standard and the
Reference solution as directed for
Procedure: the resolution,
R, between air and perflutren is not less than 2; the resolution,
R, between perflutren and decafluorobutane is not less than 5; the relative standard deviation determined from the perflutren peak response for the
Reference solution is not more than 5%; and the relative standard deviation determined from the response ratios for replicate injections of the 5% decafluorobutane5% perflutren reference standard is not more than 5%.
Procedure
Inject 20 µL of the headspace gas from the vials containing the
Reference solution and the
Test solution into the gas chromatograph. Record the chromatograms, and measure the areas for the major peaks. Calculate the quantity, in mg per mL, of perflutren in the portion of Injectable Suspension taken by the formula:
0.188M/V)(RS / RU),
in which
M is the number of µmols of decafluorobutane in an
Analysis vial after addition of the Injectable Suspension;
V is the volume, in mL, of Injectable Suspension added to the
Analysis vial; and
RS and
RU are the peak area ratios of decafluorobutane to perflutren obtained from the
Reference solution and the
Test solution, respectively. The quantity of perflutren in the Injectable Suspension is between 0.11 mg per mL and 0.33 mg per mL.
Assay for protein
Diluted antifoam reagent
Transfer 100 µL of antifoam reagent
8 to a suitable container, and dilute with water to 10 mL.
Blank preparation
Transfer 500 µL of Sodium Chloride Injection to a culture tube. Dilute the contents of the tube with water to 2 mL, and add 10 µL of Diluted antifoam reagent.
Standard preparations
Transfer 25-, 50-, 62.5-, 75-, and 100-µL aliquots of protein standard solution
9 containing 8 g per dL into separate tubes. Dilute the contents of each tube with water to 2.00 mL, and add 10 µL of the
Diluted antifoam reagent to each tube. During the
Procedure, the addition of 3.0 mL of
biuret reagent TS to each of the tubes produces
Standard preparations with protein concentrations of 0.4, 0.8, 1.0, 1.2, and 1.6 mg per mL.
Assay preparation
Equilibrate each container of Injectable Suspension to room temperature, and mix each for at least 5 minutes to ensure a homogeneous suspension. Vent the container, and transfer 500-µL aliquots into separate tubes. Dilute the contents of each tube with water to 2 mL, and add 10 µL of Diluted antifoam reagent.
Procedure
To each of the tubes containing the
Blank preparation, Standard preparations, and
Assay preparation, add 3.0 mL of
biuret reagent TS, mix, and allow to stand for 30 minutes, accurately timed, for maximum color development. The
Blank preparation, Standard preparations, and
Assay preparation are treated identically. Using the
Blank preparation, set the absorbance equal to zero. Determine the absorbance of each of the
Standard preparations and the
Assay preparation in 1-cm cells with a suitable spectrophotometer at a wavelength of 540 nm. Using linear regression, analyze the data obtained for each of the
Standard preparations. Calculate the correlation coefficient, slope, and
y-intercept values: the correlation coefficient is not less than 0.995. Calculate the quantity, in mg, of protein in each mL of the Injectable Suspension by the formula:
10[(AU y-intercept)/slope],
in which 10 is the dilution factor; and
AU is the absorbance of the
Assay preparation: the calculated quantity of protein in the Injectable Suspension is between 8 and 12 mg per mL.