206 ALUMINUM
This procedure is provided to demonstrate that the content of aluminum (Al) does not exceed the limit given in the individual monograph of a substance labeled as intended for use in hemodialysis. [NOTEThe Standard Preparations and the Test Preparation may be modified, if necessary, to obtain solutions of suitable concentrations adaptable to the linear or working range of the instrument.]
Nitric Acid Diluent
Transfer 40 mL of nitric acid to a 1000-mL volumetric flask, and dilute with water to volume.
Standard Preparations
Treat some aluminum wire with 6 N hydrochloric acid at 80
for a few minutes. Dissolve about 100 mg of the treated wire, accurately weighed, in a mixture of 10 mL of hydrochloric acid and 2 mL of nitric acid by heating at about 80
for approximately 30 minutes. Continue heating until the volume is reduced to about 4 mL. Cool to room temperature, and add 4 mL of water. Evaporate to about 2 mL by heating. Cool, and transfer this solution, with the aid of water, to a 100-mL volumetric flask, dilute with water to volume, and mix. Transfer 10.0 mL of this solution to a second 100-mL volumetric flask, dilute with water to volume, and mix. Transfer 1.0 mL of this solution to a third 100-mL volumetric flask, dilute with water to volume, and mix. The concentration of aluminum in this
Standard Preparation is about 1.0 µg per mL. If a more diluted
Standard Preparation is required, transfer 1.0-, 2.0-, and 4.0-mL portions of this solution to separate 100-mL volumetric flasks, dilute with
Nitric Acid Diluent to volume, and mix. These solutions contain 0.01, 0.02, and 0.04 µg of Al per mL, respectively.
Test Preparation
Unless otherwise directed in the monograph, transfer an accurately weighed amount (in g) of the test substance, as specified in the monograph, to a 100-mL plastic volumetric flask, add 50 mL of water, and sonicate for 30 minutes. Add 4 mL of nitric acid, dilute with water to volume, and mix.
Procedure
Determine the absorbances of the
Standard Preparations and the
Test Preparation at the aluminum emission line at 309.3 nm with a suitable atomic absorption spectrophotometer (see
Spectrophotometry and Light-Scattering 851) equipped with an aluminum hollow-cathode lamp and a flameless electrically heated furnace, using the
Nitric Acid Diluent as the blank. Plot the absorbances of the
Standard Preparations versus the content of Al, in µg per mL, drawing a straight line best fitting the three points. From the graph so obtained, determine the quantity, in µg, of Al in each mL of the
Test Preparation. Calculate the amount of Al in the specimen taken, in µg per g, by multiplying this value by 100/
W, where
W is the weight, in g, of the substance taken to prepare the
Test Preparation.