Chromatographic purity
Solution A
Prepare a filtered and degassed mixture of 0.17 M acetic acid and methanol (125:8). Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Solution B:
methanol, filtered and degassed.
Mobile phase
Use variable mixtures of
Solution A and
Solution B as directed for
Chromatographic system. Make adjustments to either solution as necessary (see
System Suitability under
Chromatography 621).
System suitability solution
Dissolve suitable quantities of purine and
USP Acyclovir RS in
Solution A to obtain a solution containing about 0.5 µg of each per mL.
Acyclovir standard solution
Dissolve an accurately weighed quantity of
USP Acyclovir RS in
Solution A, and dilute quantitatively, and stepwise if necessary, with
Solution A to obtain a solution having a known concentration of about 5 µg per mL.
Guanine solution
Dissolve about 25 mg of guanine, accurately weighed, in 50 mL of 0.1 N sodium hydroxide in a 500-mL volumetric flask, dilute with water to volume, and mix.
Standard solution 1
Transfer 5.0 mL of Acyclovir standard solution to a 50-mL volumetric flask, dilute with Solution A to volume, and mix.
Standard solution 2
Transfer 5.0 mL of Guanine solution to a 50-mL volumetric flask, dilute with Solution A to volume, and mix.
Test solution
Constitute and combine not fewer than 10 vials of Acyclovir for Injection. Transfer an accurately measured quantity, equivalent to about 100 mg of acyclovir, to a 200-mL volumetric flask, dilute with Solution A to volume, and mix.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 254-nm detector and a 4.6-mm × 25-cm column that contains packing L1. The flow rate is about 1 mL per minute. The chromatograph is programmed as follows.
Time (minutes) |
Solution A
(%) |
Solution B
(%) |
Elution |
0 |
100 |
0 |
equilibration |
015 |
100 |
0 |
isocratic |
1545 |
100®65 |
0®35 |
linear gradient |
4546 |
65®100 |
35®0 |
linear gradient |
4656 |
100 |
0 |
re-equilibration |
Chromatograph the
System suitability solution, and record the peak responses as directed for
Procedure: the resolution,
R, between purine and acyclovir is not less than 2.0. Chromatograph
Standard solution 1 and
Standard solution 2, and record the peak responses as directed for
Procedure: the typical retention times for guanine and acyclovir are about 5.8 minutes and 14 minutes, respectively; and the relative standard deviation of the acyclovir peak area and the guanine peak area for replicate injections is not more than 1%.
Procedure
Separately inject equal volumes (about 50 µL) of the
Standard solution and the
Test solution into the chromatograph, record the chromatograms, and measure the peak area responses. Calculate the percentage of guanine in the Acyclovir for Injection by the formula:
20,000(C/W)(rg /rsg),
in which
C is the concentration, in mg per mL, of guanine in the
Standard solution; W is the total weight, in mg, of acyclovir in the
Test solution based on the label claim;
rg is the peak response for guanine, if present, in the
Test solution; and
rsg is the peak response of guanine in the
Standard solution: not more than 1.0% is found. Calculate the percentage of each other impurity in the Acyclovir for Injection by the formula:
20,000(C/W)(ri / rS),
in which
C is the concentration, in mg per mL, of
USP Acyclovir RS in the
Standard solution; W is the total weight, in mg, of acyclovir in the
Test solution based on the label claim;
ri is the peak response for each impurity; and
rS is the peak response of acyclovir in the
Standard solution: not more than 0.15% for any peak having a relative retention time of about 0.7 compared to the acyclovir peak is found; not more than 0.5% of any other individual impurity is found; and the total of all other impurities is not more than 1.0%.
Assay
Mobile phase
Prepare a filtered and degassed solution of 0.02 M acetic acid. Make adjustments if necessary (see
System Suitability under
Chromatography 621).
System suitability preparation 1
Dissolve accurately weighed quantities of
USP Acyclovir RS and guanine in 0.1 N sodium hydroxide, and dilute quantitatively, and stepwise if necessary, with 0.1 N sodium hydroxide to obtain a solution having known concentrations of about 0.1 mg of each per mL.
System suitability preparation 2
Dissolve an accurately weighed quantity of guanine in 0.1 N sodium hydroxide, and dilute quantitatively, and stepwise if necessary, with 0.1 N sodium hydroxide to obtain a solution having a known concentration of about 2.0 µg per mL.
Standard preparation
Dissolve accurately weighed quantities of
USP Acyclovir RS in 0.1 N sodium hydroxide, and dilute quantitatively, and stepwise if necessary, with 0.1 N sodium hydroxide to obtain a solution having a known concentration of about 0.1 mg per mL.
Assay preparation
Constitute, with water, 1 vial of Acyclovir for Injection. Transfer an accurately weighed amount of this solution, equivalent to about 10 mg of acyclovir, to a 100-mL volumetric flask, and dilute with water to volume.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 254-nm detector and a 4.2-mm × 25-cm column that contains packing L1. The flow rate is about 1.5 mL per minute. Chromatograph
System suitability preparation 1, and record the peak responses for acyclovir as directed for
Procedure: the relative retention times are about 0.6 for guanine and 1.0 for acyclovir; the resolution,
R, between guanine and acyclovir is not less than 2.0; and the relative standard deviation for replicate injections for the acyclovir peak is not more than 2.0%. Chromatograph
System suitability preparation 2, and record the peak responses as directed for
Procedure: the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
Separately inject equal volumes (about 20 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the peak responses. Calculate the quantity, in mg, of acyclovir (C
8H
11N
5O
3) in the portion of Acyclovir for Injection taken by the formula:
100C(rU / rS),
in which
C is the concentration, in mg per mL, of
USP Acyclovir RS in the
Standard preparation; and
rU and
rS are the peak responses obtained from the
Assay preparation and the
Standard preparation, respectively.