Chromatographic purity
Test solution
Prepare a solution of Triazolam in chloroform containing about 2 mg per mL.
Chromatographic system (see Chromatography 621)
The gas chromatograph is equipped with a flame-ionization detector, and contains a 3-mm × 120-cm glass column packed with 3% phase G6 on support S1AB. The column and injection port are maintained at a temperature of about 240
. The detector is maintained at a temperature of about 20
to 50
above column temperature. The carrier gas is helium.
Procedure
[NOTEAllow about three times the elution time of the major component before making another injection.
] Chromatograph about 4 µL of the
Test solution. Calculate the total percentage of impurities taken by the formula:
100S/(S + A),
in which
S is the sum of the areas of each of the minor component peaks detected, and
A is the area of the major component peak. The total amount of impurities present is not more than 1.5%.
Assay
Mobile phase
Prepare a filtered and degassed mixture of acetonitrile, chloroform, butyl alcohol, water, and glacial acetic acid (850:80:50:20:0.5). Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Internal standard solution
Dissolve an accurately weighed quantity of alprazolam in acetonitrile, and dilute with acetonitrile to obtain a solution having a known concentration of about 0.3 mg per mL.
Standard stock solution
Dissolve an accurately weighed quantity of
USP Triazolam RS in
Internal standard solution, and dilute with
Internal standard solution to obtain a solution having a known concentration of about 0.25 mg per mL.
Assay stock solution
Transfer about 2.5 mg of Triazolam, accurately weighed, to a 10-mL volumetric flask, dissolve in and dilute with Internal standard solution to volume, and mix.
Standard preparation
Dilute an accurately measured portion of the
Standard stock solution with acetonitrile to obtain a solution having a known concentration of
USP Triazolam RS of about 0.025 mg per mL.
Assay preparation
Transfer about 5 mL of the Assay stock solution, accurately measured, to a 50-mL volumetric flask, dilute with acetonitrile to volume, and mix.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 254-nm detector and a 4.6-mm × 30-cm column that contains packing L3. The flow rate is about 2.0 mL per minute. Chromatograph the
Standard preparation, and record the peak responses as directed under
Procedure: the relative retention times are 1 for triazolam and about 1.4 for the internal standard, the resolution,
R, between the internal standard and triazolam is not less than 2.0, and the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
Separately inject equal volumes (about 20 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the area responses for the major peaks. Calculate the quantity, in mg, of C
17H
12Cl
2N
4 in the portion of Triazolam taken by the formula:
CV(RU / RS),
in which
C is the concentration, in mg per mL, of
USP Triazolam RS in the
Standard stock solution; V is the volume of internal standard used to prepare the
Assay stock solution; and
RU and
RS are the ratios of the internal standard peak area to the alprazolam peak area obtained from the
Assay preparation and the
Standard preparation, respectively.