Chromatographic purity
Triethylamine phosphate solution, Mobile phase, System suitability solution, and Chromatographic system
Proceed as directed in the Assay.
Standard solution
Use the Standard preparation, prepared as directed in the Assay.
Test solution
Use the Assay preparation.
Procedure
Separately inject equal volumes (about 20 µL) of the
Standard solution and the
Test solution into the chromatograph, record the chromatograms, and measure the peak responses. Calculate the percentage of each impurity in the portion of Saquinavir Mesylate taken by the formula:
100F(CS /CU)(ri / rS),
in which
F is a response factor and is equal to 2 for peaks, if present, at a retention time of 0.32 relative to saquinavir, to 0.5 for peaks, if present, at retention times of about 0.38 and 0.53 relative to saquinavir, and to 1 for all other peaks;
CS is the concentration, in mg per mL, of
USP Saquinavir Mesylate RS in the Standard solution;
CU is the concentration, in mg per mL, of Saquinavir Mesylate in the
Test solution; ri is the peak response for each impurity obtained from the
Test solution; and
rS is the peak response for saquinavir obtained from the
Standard solution: not more than 0.1% of any individual impurity is found; and not more than 0.5% of total impurities is found.
Assay
Triethylamine phosphate solution
Transfer 10 mL of triethylamine to a 1-liter volumetric flask, dilute with water to volume, and mix. Adjust with phosphoric acid to a pH of 2.5, and filter.
Mobile phase
Prepare a filtered and degassed mixture of
Triethylamine phosphate solution, tetrahydrofuran, and acetonitrile (14:5:1).
[NOTEProtect from light.
] Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Standard preparation
Dissolve an accurately weighed quantity of
USP Saquinavir Mesylate RS in
Mobile phase, and dilute quantitatively, and stepwise if necessary, with
Mobile phase to obtain a solution having a known concentration of about 0.25 mg per mL.
Assay preparation
Transfer about 12.5 mg of Saquinavir Mesylate, accurately weighed, to a 50-mL volumetric flask, dissolve in and dilute with Mobile phase to volume, and mix for about 20 minutes.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 210-nm detector and a 4.6-mm × 25-cm column that contains packing L1. The column temperature is maintained at 20
, and the flow rate is about 1 mL per minute. Chromatograph the
System suitability solution, and record the peak responses as directed for
Procedure: the relative retention times are about 0.89 for saquinavir related compound A and 1.0 for saquinavir; and the resolution,
R, between saquinavir related compound A and saquinavir is not less than 1.5. Chromatograph the
Standard preparation, and record the peak responses as directed for
Procedure: the column efficiency is not less than 500 theoretical plates; and the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
Separately inject equal volumes (about 20 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in mg, of C
38H
50N
6O
5·CH
4O
3S in the portion of Saquinavir Mesylate taken by the formula:
50C(rU / rS),
in which
C is the concentration, in mg per mL, of
USP Saquinavir Mesylate RS in the
Standard preparation; and
rU and
rS are the peak responses obtained from the
Assay preparation and the
Standard preparation, respectively.