Chromatographic system (see Chromatography 621)
The gas chromatograph is equipped with a flame-ionization detector maintained at about 275
, a split injection system with a split ratio of about 30:1, and a 0.32-mm × 30-m column bonded with a 0.5-µm layer of phase G42. The column temperature is programmed to be maintained at about 80
for 0.5 minute, then to increase to about 220
at a rate of 20
per minute, and to hold at about 220
for 10 minutes. The injection port temperature is maintained at about 85
. Helium is used as the carrier gas at a flow rate of about 2.3 mL per minute. Chromatograph the
System suitability solution, and record the peak responses as directed for
Procedure: the relative retention times are about 0.9 for tributyl citrate and 1.0 for acetyltributyl citrate; the resolution,
R, between tributyl citrate and acetyltributyl citrate is not less than 1.5; and the relative standard deviation for replicate injections is not more than 2.0% determined from both the tributyl citrate and acetyltributyl citrate peaks, based on area percent calculation.
Procedure
Inject 1 µL of the
Assay preparation into the chromatograph, record the chromatogram, and measure all of the peak areas, excluding the solvent peak. Calculate the percentage of C
20H
34O
8 in the portion of Acetyltributyl Citrate taken by the formula:
100(A/B),
in which
A is the acetyltributyl citrate peak area response; and
B is the sum of the area responses of all the peaks.