Chromatographic purity
Mobile phase and Chromatographic system
Prepare as directed in the Assay.
Test solution
Transfer about 10 mg of Atenolol to a 100-mL volumetric flask, dissolve in and dilute with Mobile phase to volume, and mix.
Diluted test solution
Transfer 0.50 mL of the Test solution to a 100-mL volumetric flask, dilute with Mobile phase to volume, and mix.
Procedure
Separately inject equal volumes (about 50 µL) of the
Test solution and the
Diluted test solution into the chromatograph, record the chromatograms, and measure the areas for all the peaks.
[NOTEChromatograph the
Test solution for a period of time 6 times the retention time of the atenolol peak.
] Calculate the percentage of each impurity observed in the chromatogram obtained from the
Test solution by the formula:
0.5(ri / rA),
in which
ri is the peak response of an individual impurity in the chromatogram obtained from the
Test solution; and
rA is the response of the main atenolol peak in the chromatogram obtained from the
Diluted test solution. Not more than 0.25% of any individual impurity is found, and the sum of all the impurities is not more than 0.5%.
Assay
Mobile phase
Dissolve 1.1 g of sodium 1-heptanesulfonate and 0.71 g of anhydrous dibasic sodium phosphate in 700 mL of water. Add 2 mL of dibutylamine, and adjust with 0.8
M phosphoric acid to a pH of 3.0. Add 300 mL of methanol, mix, and pass through a filter having a 0.5 µm or finer porosity. Degas this solution before use. Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Standard preparation
Quantitatively dissolve an accurately weighed quantity of
USP Atenolol RS in
Mobile phase to obtain a solution having a known concentration of about 0.01 mg per mL.
Assay preparation
Transfer about 100 mg of Atenolol, accurately weighed, to a 100-mL volumetric flask, add 50 mL of Mobile phase, and sonicate for about 5 minutes. Dilute with Mobile phase to volume, and mix. Transfer 5.0 mL of this solution to a 50-mL volumetric flask, dilute with Mobile phase to volume, and mix. Transfer 5.0 mL of this solution to a second 50-mL volumetric flask, dilute with Mobile phase to volume, and mix.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 226-nm detector and a 3.9-mm × 30-cm column that contains packing L1. The flow rate is about 0.6 mL per minute. Chromatograph the
Standard preparation, and record the peak responses as directed for
Procedure: the column efficiency is not less than 5000 theoretical plates; the tailing factor is not more than 2.0; and the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
Separately inject equal volumes (about 10 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in mg, of C
14H
22N
2O
3 in the portion of Atenolol taken by the formula:
100C(rU / rS),
in which
C is the concentration, in mg per mL, of
USP Atenolol RS in the
Standard preparation; and
rU and
rS are the atenolol peak responses obtained from the
Assay preparation and the
Standard preparation, respectively.