Packaging and storage
Preserve in single-dose or in multiple-dose containers, preferably of Type I glass, at a temperature between 2
and 30
, protected from light.
Limit of ondansetron related compound D
Mobile phase, Standard solution, System suitability solution, and Chromatographic system
Proceed as directed in the test for
Limit of ondansetron related compound D under
Ondansetron Hydrochloride.
Test solution
Transfer an accurately measured volume of Injection, equivalent to about 10 mg of ondansetron, to a 25-mL volumetric flask, dilute with Mobile phase to volume, and mix.
Procedure
Separately inject equal volumes (about 20 µL) of the
Standard solution and the
Test solution into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity of ondansetron related compound D in the volume of Injection taken by the formula:
(2.5 / V)(CS / CA)(rU / rS),
in which
V is the volume, in mL, of Injection taken;
CS is the concentration, in µg per mL, of ondansetron related compound D in the
Standard preparation; CA is the concentration, in mg per mL, of ondansetron in the Injection, as determined in the
Assay; and
rU and
rS are the peak responses obtained from the
Test preparation and the
Standard preparation, respectively: not more than 0.12% is found.
Chromatographic purity
Mobile phase and Chromatographic system
Proceed as directed in the Assay.
Standard solution
Proceed as directed for Standard preparation in the Assay.
Test solution
Use the Assay preparation.
Procedure
Inject a volume (about 10 µL) of the
Test solution into the chromatograph, record the chromatogram, and measure the peak responses. Calculate the percentage of each impurity in the volume of Injection taken by the formula:
100(ri / rs),
in which
ri is the peak response for each impurity; and
rs is the sum of the responses of all of the peaks: not more than 0.2% of any individual impurity is found, and the total of all impurities is not more than 0.5%.
Assay
Mobile phase
Prepare a filtered and degassed mixture of 0.02 M monobasic potassium phosphate (previously adjusted with 1 M sodium hydroxide to a pH of 5.4), and acetonitrile (50:50). Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Standard preparation
Dissolve an accurately weighed quantity of
USP Ondansetron Hydrochloride RS in
Mobile phase, and dilute quantitatively, and stepwise if necessary, with
Mobile phase to obtain a solution having a known concentration of about 0.1 mg per mL.
Assay preparation
Transfer an accurately measured volume of Injection, equivalent to about 2 mg of ondansetron, to a 25-mL volumetric flask, dilute with Mobile phase to volume, and mix.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 216-nm detector and a 4.6-mm × 20-cm column that contains packing L10. The flow rate is about 1.5 mL per minute. Chromatograph the
System suitability solution, and record the peak responses as directed for
Procedure: the relative retention times are about 1.0 for ondansetron and 1.1 for ondansetron related compound A; and the resolution,
R, between ondansetron related compound A and ondansetron is not less than 1.5. Chromatograph the
Standard preparation, and record the peak responses as directed for
Procedure: the tailing factor is not more than 2.0; and the relative standard deviation for replicate injections is not more than 1.5%.
Procedure
Separately inject equal volumes (about 10 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in mg, of ondansetron (C
18H
19N
3O) in each mL of the Injection taken by the formula:
(293.36 / 329.82)(25C / V)(rU / rS),
in which 293.36 and 329.82 are the molecular weights of ondansetron and anhydrous ondansetron hydrochloride, respectively;
C is the concentration, in mg per mL, on the anhydrous basis, of
USP Ondansetron Hydrochloride RS in the
Standard preparation; V is the volume, in mL, of Injection taken; and
rU and
rS are the peak responses obtained from the
Assay preparation and the
Standard preparation, respectively.