Dissolution 711
Test 1
Medium:
0.5% sodium lauryl sulfate in water; 900 mL.
Apparatus 2:
25 rpm.
Time:
30 minutes.
Standard solution
Transfer about 45 mg, accurately weighed, of
USP Megestrol Acetate RS to a 250-mL volumetric flask, add about 12 mL of methanol, and put the flask in a warm water bath until the solid is dissolved. Dilute with
Medium to volume. The final concentration is about 18 µg of megestrol acetate per mL.
Procedure
Transfer to the surface of the
Medium in the dissolution vessel an accurately measured volume of Oral Suspension, freshly mixed and free from air bubbles, equivalent to about 160 mg of megestrol acetate. Determine the amount of C
24H
32O
4 dissolved by employing UV absorption at the wavelength of maximum absorbance at about 292 nm on filtered portions of the solution under test, in comparison with the
Standard solution. Calculate the percentage of megestrol acetate (C
24H
32O
4) released by the formula:
in which
AU and
AS are the absorbances obtained from the solution under test and the
Standard solution, respectively;
CS is the concentration, in mg per mL, of the
Standard solution; V is the sample volume, in mL, of Oral Suspension taken; 900 is the volume, in mL, of
Medium; 100 is the conversion factor to percentage; and
LC is the label claim, in mg per mL.
Tolerances
Not less than 80% (Q) of the labeled amount of C24H32O4 is dissolved in 30 minutes.
Test 2
If the product complies with this test, the labeling indicates that it meets USP Dissolution Test 2.
Medium:
0.5% sodium lauryl sulfate in water; 900 mL.
Apparatus 2:
25 rpm.
Time:
30 minutes.
Standard solution
Transfer about 45 mg, accurately weighed, of
USP Megestrol Acetate RS to a 250-mL volumetric flask. Add about 5 mL of methanol, and mix. Dilute with
Medium to volume. Transfer 10 mL of this solution to a 100-mL volumetric flask, and dilute with
Medium to volume. The final concentration is about 18 µg per mL.
Test solution
[NOTEUse a separate syringe for each vessel.] Withdraw more than 10 mL of the Oral Suspension, using a 10-mL syringe with a long cannula. Remove air bubbles from the syringe. Adjust the volume to the 10-mL mark on the syringe, and remove the needle. Wipe the tip of the syringe, and accurately weigh (gross weight). Operate the apparatus, and rapidly dispense the Oral Suspension to the side of the vessel at about halfway from the bottom. Similarly dispense the Oral Suspension into other vessels. Accurately weigh each syringe after dispensing the sample (tare weight). Record sample weights. After completion of the dissolution, pass an aliquot through a nylon filter having a 0.45-µm porosity, and dilute 2.0 mL of the filtrate with Medium to 50.0 mL to obtain a solution having a theoretical concentration of about 18 µg per mL.
Procedure
Determine the amount of C
24H
32O
4 dissolved by employing UV absorption at the wavelength of maximum absorbance at about 292 nm, using 0.5-cm pathlength cuvettes, on the
Test solution in comparison with the
Standard solution. Calculate the percentage of megestrol acetate (C
24H
32O
4) released by the formula:
in which
AU and
AS are the absorbances obtained from the
Test solution and the
Standard solution, respectively;
CS is the concentration, in mg per mL, of the
Standard solution; d is the density, in mg per mL, of the Oral Suspension obtained by dividing the weight of Oral Suspension taken by 10 mL;
WU is the weight, in mg, of Oral Suspension taken; 900 is the volume, in mL, of
Medium; 100 is the conversion factor to percentage; and
LC is the label claim, in mg per mL.
Tolerances
Not less than 80% (Q) of the labeled amount of C24H32O4 is dissolved in 30 minutes.
Test 3
If the product complies with this test, the labeling indicates that it meets USP Dissolution Test 3.
Medium:
0.5% sodium lauryl sulfate in degassed water; 900 mL. Use ultrapure sodium lauryl sulfate with an assay content of not less than 99.0%.
Apparatus 2:
50 rpm.
Time:
30 minutes.
Determine the amount of C24H32O4 dissolved by employing the following method.
Mobile phase
Proceed as directed in the Assay.
Standard solution
Transfer about 11.5 mg, accurately weighed, of
USP Megestrol Acetate RS to a 25-mL volumetric flask, and dilute with
Mobile phase to volume.
Test solution
Proceed as directed for Test 2, introducing the sample into the vessel over a 10- to 15-second period (about 1 mL per second).
Procedure
Separately inject equal volumes (about 10 µL) of the
Standard solution and the
Test solution into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the percentage of megestrol acetate (C
24H
32O
4) released by the formula:
in which
rU and
rS are the peak responses obtained from the
Test solution and the
Standard solution, respectively;
CS is the concentration, in mg per mL, of the
Standard solution; d is the density, in mg per mL, of the Oral Suspension obtained by dividing the weight of Oral Suspension taken by 10 mL;
WU is the weight, in mg, of Oral Suspension taken; 900 is the volume, in mL, of
Medium; 100 is the conversion factor to percentage; and
LC is the label claim, in mg per mL.
Tolerances
Not less than 80% (Q) of the labeled amount of C24H32O4 is dissolved in 30 minutes.
Assay
Mobile phase
Prepare a filtered and degassed mixture of acetonitrile and water (11:9). Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Standard preparation
Dissolve an accurately weighed quantity of Megestrol Acetate RS in Mobile phase, and dilute quantitatively, and stepwise if necessary, with Mobile phase to obtain a solution having a known concentration of about 80 µg per mL.
Assay preparation
Transfer an accurately measured volume of Oral Suspension, equivalent to about 160 mg of megestrol acetate, to a 100-mL volumetric flask, and dissolve in and dilute with Mobile phase to volume. Transfer 5.0 mL of the solution so obtained to a 100-mL volumetric flask, dilute with Mobile phase to volume, and mix.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 280-nm detector and a 3.9-mm × 30-cm column that contains packing L1. The flow rate is about 1.5 mL per minute. Chromatograph the
Standard preparation, and record the peak responses as directed for
Procedure: the column efficiency is not less than 2500 theoretical plates; the tailing factor is not more than 1.4; and the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
Separately inject equal volumes (about 25 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in mg, of megestrol acetate (C
24H
32O
4) in the portion of Oral Suspension taken by the formula:
2C(rU / rS),
in which
C is the concentration, in µg per mL, of
USP Megestrol Acetate RS in the
Standard preparation; and
rU and
rS are the peak responses obtained from the
Assay preparation and the
Standard preparation, respectively.