Identification
A:
Dissolve about 100 mg in 5 mL of water, add 5 mL of 1 N sodium hydroxide, cool to about 10
, add 1 mL of a mixture of absolute ether and benzoyl chloride (2:1), insert the stopper, and shake for 3 minutes. Filter the precipitate, wash with about 10 mL of cold water, and recrystallize from diluted alcohol: the crystals of the benzoyl derivative of amphetamine so obtained, after drying at 80
for 3 hours, melt between 131
and 135
, the procedure for
Class I being used (see
Melting Range or Temperature 741).
B:
A solution (1 in 10) responds to the tests for
Sulfate 191.
Chromatographic purity
Diluent
Dilute 3.12 mL of phosphoric acid with water to 1000 mL.
Buffer solution
Dissolve 2.16 g of sodium 1-octanesulfonate in 1000 mL of water, and add 1.0 mL of triethylamine. Mix, and adjust with phosphoric acid to a pH of 2.5.
Mobile phase
Prepare a filtered and degassed mixture of
Buffer solution, acetonitrile, and methanol (144:37:19). Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Standard stock solution
Dissolve an accurately weighed quantity of
USP Dextroamphetamine Sulfate RS in
Diluent to obtain a solution having a known concentration of about 0.3 mg per mL.
Standard solution
Dilute an accurately measured volume of Standard stock solution in Diluent to obtain a solution having a known concentration of about 0.003 mg per mL.
Test solution
Transfer about 30 mg of Amphetamine Sulfate, accurately weighed, to a 100-mL volumetric flask. Dissolve in 50 mL of Diluent, sonicate for 5 minutes, dilute with Diluent to volume, and mix.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 215-nm detector and a 4.6-mm × 15-cm column that contains 5-µm packing L1. The flow rate is about 1 mL per minute. Chromatograph the
Standard stock solution, and record the peak responses as directed for
Procedure: the tailing factor is not more than 2.0; and the relative standard deviation for replicate injections is not more than 2.0%. Chromatograph the
Test solution, and record the peak responses as directed for
Procedure: the resolution,
R, between amphetamine and any adjacent peak, if any, is not less than 1.5.
Procedure
Separately inject equal volumes (about 50 µL) of the
Standard solution and the
Test solution into the chromatograph, record the chromatograms, and measure the peak responses. Calculate the percentage of each impurity in the portion of Amphetamine Sulfate taken by the formula:
10,000(C/W)(ri / rS),
in which
C is the concentration, in mg per mL, of
USP Dextroamphetamine Sulfate RS in the
Standard solution; W is the weight, in mg, of Amphetamine Sulfate taken to prepare the
Test solution; ri is the peak response for each impurity obtained from the
Test solution; and
rS is the peak response for amphetamine obtained from the
Standard solution: not more than 0.1% of any individual impurity is found, and not more than 0.5% of total impurities is found.
Change to read:
Assay
Dissolve about 500 mg of Amphetamine Sulfate, accurately weighed, in 50 mL of glacial acetic acid, and titrate with 0.1 N perchloric acid VS, determining the endpoint potentiometrically. Perform a blank determination, and make any necessary correction (see
Titrimetry 541). Each mL of 0.1 N perchloric acid is equivalent to 36.85 mg of (C
9H
13N)
2·H
2SO
4.
USP29