Content of (Z)-isomer
Mobile phase, Standard preparation, System suitability solution, and Chromatographic system
Proceed as directed in the Assay.
Test solution
Use the Assay preparation prepared as directed in the Assay.
Procedure
Inject a volume (about 50 µL) of the
Test solution into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the percentage of (
Z)-isomer in the portion of Clomiphene Citrate taken by the formula:
100(rZ / ri),
in which
rZ is the peak response of the (
Z)-isomer in the
Test solution, and
ri is the sum of all of the peak responses obtained from the
Test solution.
Related compounds
Mobile phase and System suitability solution
Proceed as directed in the Assay.
Standard solution
Dissolve an accurately weighed quantity of
USP Clomiphene Citrate RS in
Mobile phase, and dilute quantitatively, and stepwise if necessary, with
Mobile phase to obtain a solution having a known concentration of about 1.0 µg per mL.
Test solution
Use the Assay preparation.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 290-nm detector and a 4.6-mm × 25-cm column that contains packing L26. The flow rate is about 1 mL per minute. Chromatograph the
System suitability solution, and record the peak responses as directed for
Procedure: the relative retention times are about 0.9 for clomiphene related compound A, 1.0 for the (
Z)-isomer, and 1.2 for the (
E)-isomer; the resolution,
R, between clomiphene related compound A and the (
Z)-isomer is not less than 1.0; and the resolution,
R, between the (
Z)-isomer and the (
E)-isomer is not less than 1.5. Chromatograph the
Standard solution, and record the peak responses as directed for
Procedure: the column efficiency is not less than 2000 theoretical plates for the (
E)-isomer; the tailing factor is not more than 3.0 for the (
E)-isomer; and the relative standard deviation for replicate injections is not more than 2.0% determined from both the (
E)- and (
Z)-isomers.
Procedure
Inject a volume (about 50 µL) of the
Test solution into the chromatograph, record the chromatogram, and measure the peak responses. Calculate the percentage of each impurity in the portion of Clomiphene Citrate taken by the formula:
100(ri / rs),
in which
ri is the peak response for each impurity; and
rs is the sum of the responses of all of the peaks: not more than 2.0% of clomiphene related compound A is found; not more than 0.5% of any other individual impurity is found; and not more than 2.0% of total impurities is found.
Assay
Mobile phase
Prepare a filtered and degassed mixture of methanol, water, and triethylamine (55:45:0.3). Adjust with phosphoric acid to a pH of 2.5. Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Standard preparation
[NOTEUse actinic glassware for all solutions.
] Dissolve an accurately weighed quantity of
USP Clomiphene Citrate RS in
Mobile phase, and dilute quantitatively, and stepwise if necessary, with
Mobile phase to obtain a solution having a known concentration of about 0.05 mg per mL.
Assay preparation
Transfer about 50 mg of the Clomiphene Citrate, accurately weighed, to a 100-mL volumetric flask, dilute with Mobile phase to volume, mix, and filter. Transfer 10.0 mL of this solution to a 100-mL volumetric flask, dilute with Mobile phase to volume, and mix.
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 233-nm detector and a 4.6-mm × 25-cm column that contains packing L26. The flow rate is about 1 mL per minute. Chromatograph the
System suitability solution, and record the peak responses as directed for
Procedure: the relative retention times are about 0.9 for clomiphene related compound A, 1.0 for (
Z)-isomer, and 1.2 for (
E)-isomer; and the resolution,
R, between clomiphene related compound A and (
Z)-isomer is not less than 1.0 and between (
Z)-isomer and (
E)-isomer is not less than 1.5. Chromatograph the
Standard preparation, and record the peak responses as directed for
Procedure: the column efficiency is not less than 2000 theoretical plates for the (
E)-isomer; the tailing factor is not more than 3.0 for the (
E)-isomer; and the relative standard deviation for replicate injections is not more than 2.0% for both (
E)- and (
Z)-isomers.
Procedure
Separately inject equal volumes (about 50 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in mg, of C
26H
28ClNO·C
6H
8O
7 in the portion of Clomiphene Citrate taken by the formula:
1000C(rUE + rUZ / rSE + rSZ),
in which
C is the concentration, in mg per mL, of
USP Clomiphene Citrate RS in the
Standard preparation; rUE and
rUZ are the peak responses obtained from the
Assay preparation for the (
E)- and (
Z)-isomers, respectively; and
rSE and
rSZ are the peak responses obtained from the
Standard preparation for the (
E)- and (
Z)-isomers, respectively.