Chromatographic system (see Chromatography 621)
The gas chromatograph is equipped with a flame-ionization detector and a 4-mm × 1.8-m column packed with 3% phase G16 on support S1A. The carrier gas is dry nitrogen, flowing at a rate of about 30 mL per minute. The column temperature is maintained at 180
, and the injection port and detector block temperatures are maintained at 200
. Chromatograph the
Standard solution, and record the peak responses as directed for
Procedure: the resolution,
R, between 3,5-dimethylphenol and chloroxylenol related compound A is not less than 4.5; and the relative standard deviation for replicate injections is not more than 10%.
Procedure
Separately inject equal volumes (about 1 µL) of the
Standard solution and the
Test solution into the chromatograph, record the chromatograms, and measure the areas for all the peaks. Calculate the percentages of 3,5-dimethylphenol (C
8H
10O) and chloroxylenol related compound A (C
8H
9ClO) in the portion of Chloroxylenol taken by the formula:
0.2(rU / rS),
in which
rU and
rS are the peak areas of the appropriate analyte obtained from the
Test solution and the
Standard solution, respectively: not more than 0.2% of 3,5-dimethylphenol or chloroxylenol related compound A is found. Calculate the percentage of each impurity in the portion of Chloroxylenol taken by the formula:
100(ri / rs),
in which
ri is the area of each peak obtained from the
Test solution, excluding that of the main chloroxylenol peak, the 3,5-dimethylphenol peak, and the chloroxylenol related compound A peak; and
rs is the sum of the areas of all the peaks: not more than 0.5% of any individual impurity is found. Calculate the percentage of total impurities in the portion of Chloroxylenol taken by the formula:
100(rT / rs),
in which
rT is the sum of the areas of all the peaks obtained from the
Test solution, excluding that of the main chloroxylenol peak; and
rs is the sum of the areas of all the peaks obtained from the
Test solution: not more than 1.5% of total impurities is found.