Assay
Internal standard solution
Dissolve about 500 mg of 4-tert-butylphenol, accurately weighed, in acetone in a 100-mL volumetric flask, add acetone to volume, and mix.
Standard preparation
Dissolve accurately weighed quantities of USP 3-tert-Butyl-4-hydroxyanisole RS and USP 2-tert-Butyl-4-hydroxyanisole RS in Internal standard solution to obtain a solution having known concentrations of about 9 mg per mL and about 1 mg per mL, respectively.
Assay preparation
Dissolve about 100 mg of Butylated Hydroxyanisole, accurately weighed, in Internal standard solution in a 10-mL volumetric flask, dilute with Internal standard solution to volume, and mix.
Chromatographic system (see Chromatography 621)
The gas chromatograph is equipped with a flame-ionization detector, and contains a 2-mm × 1.8-m stainless-steel column packed with 10% liquid phase G26 on support S1A. The carrier gas is helium. The column is maintained isothermally at a temperature between 175
and 185
. The injection port temperature is maintained at 200
, and the detector temperature is maintained at about 250
. Chromatograph the
Standard preparation, and record the peak responses as directed for
Procedure: the resolution,
R, between the 3-
tert-butyl-4-hydroxyanisole isomer and the 2-
tert-butyl-4-hydroxyanisole isomer is not less than 1.3; the tailing factor is not more than 2.0; and the relative standard deviation is not more than 2.0% determined from the 3-
tert-butyl-4-hydroxyanisole isomer and not more than 6.0% determined from the 2-
tert-butyl-4-hydroxyanisole isomer.
Procedure
Separately inject equal volumes (about 5 µL) of the
Standard preparation and the
Assay preparation into the gas chromatograph, record the chromatograms, and measure the peak areas for each isomer and the internal standard. Calculate the quantity, in mg, of each isomer in the portion of Butylated Hydroxyanisole taken by the formula:
10C(RU / RS),
in which
C is the concentration, in mg per mL, of the appropriate USP Reference Standard; and
RU and
RS are the peak area ratios of the corresponding isomer to the internal standard obtained from the
Assay preparation and the
Standard preparation, respectively. Calculate the weight, in mg, of C
11H
16O
2 in the portion of Butylated Hydroxyanisole taken by adding the quantities of the two isomers.