Lead
[
NOTEUse acid-cleaned (mixture of 5% nitric acid and 5% hydrochloric acid followed by rinsing with water) autosampler cups and volumetric glassware to avoid contamination. For the preparation of all aqueous solutions and for the rinsing of glassware before use, employ water that has been passed through a strong-acid, strong-base, mixed-bed ion-exchange resin. Select all reagents to have as low a content of lead as practicable. Store standards and samples in acid-cleaned polyethylene containers.]
Diluent
Transfer 2 mL of lead-free nitric acid into a 1000-mL volumetric flask, dilute with water to volume, and mix.
Standard stock solution
Dissolve 79.9 mg of lead nitrate in 100 mL of Diluent in a 500-mL volumetric flask, and dilute with Diluent to volume. Transfer 10.0 mL of the resulting solution into a 100-mL volumetric flask, and dilute with Diluent to volume. Each mL of the Standard stock solution contains the equivalent of 10 µg of lead.
Standard solutions
Dilute aliquots of the Standard stock solution quantitatively, and stepwise if necessary, with Diluent to obtain solutions having concentrations of 0.03 and 0.015 µg per mL.
Test solution
Transfer 160 mg of Neotame, accurately weighed, to a 10-mL volumetric flask. Dissolve in and dilute with Diluent to volume.
Blank
Use the Diluent as the blank.
Procedure
Separately inject equal volumes (about 15 µL) of the
Standard solutions, the
Blank, and the
Test solution into a suitable graphite furnace atomic absorption spectrophotometer set at 283.3 nm and equipped with an autosampler, pyrolytically coated graphite tubes, a solid pyrolytic graphite platform, and an adequate means of background correction. Use a hollow-cathode lamp as the source, use argon as the purge gas, and breathing-quality air as the alternate gas. Optimize the instrument program as recommended by the manufacturer for lead, using a char temperature of 500
and an atomization temperature of 2000
. Correct the area responses of all
Test solutions and
Standard solutions for the
Blank area response. Generate the appropriate lead calibration algorithm, and determine the lead concentration in the
Test solution, in µg per mL. Calculate the percentage of lead in the portion of Neotame taken by the formula:
1000(C/W)
in which
C is the blank-corrected lead concentration in the
Test solution, in µg per mL; and
W is the weight, in µg, of Neotame taken to prepare the
Test solution: not more than 0.0002% is found.
Related compounds
Mobile phase
Proceed as directed in the Assay.
Related compound A standard solution
Dissolve an accurately weighed quantity of USP Neotame Related Compound A RS in Mobile phase to obtain a solution having a known concentration of about 0.03 mg per mL.
Detector sensitivity solution
Transfer 2 mL of the Related compound A standard solution to a 50-mL volumetric flask, dilute with Mobile phase to volume, and mix.
Standard solution
Prepare as directed for Standard preparation in the Assay.
Test solution
Transfer about 100 mg of Neotame, accurately weighed, to a 50-mL volumetric flask, dissolve in and dilute with
Mobile phase to volume, and mix. [
NOTEThis solution is stable for up to 32 hours when stored at a temperature of 0
to 10
.]
Chromatographic system (see Chromatography 621)
Use the same system as directed in the
Assay except to chromatograph the
Related compound A standard solution and the
Detector sensitivity solution, and record the peak responses as directed for
Procedure: the signal-to-noise ratio for the
Detector sensitivity solution is not less than 10; and the relative standard deviation for replicate injections of the
Related compound A standard solution is not more than 5.0%.
Procedure
Separately inject equal volumes (about 25 µL) of the
Related compound A standard solution, the
Standard solution, and the
Test solution into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the percentage of related compound A in the portion of Neotame taken by the formula:
100(CS / CU)(rU / rS)
in which
CS is the concentration, in mg per mL, of USP Neotame Related Compound A RS in the
Related compound A standard solution; CU is the concentration, in mg per mL, of Neotame in the
Test solution; and
rU and
rS are related compound A peak responses obtained from the the
Test solution and
Related compound A standard solution, respectively: not more than 1.5% is found. Calculate the percentage of other impurites in the portion of Neotame taken by the formula:
100(CS / CU)(Ai / AS)
in which
CS is the concentration, in mg per mL, of USP Neotame RS in the
Standard solution; and
CU is the concentration, in mg per mL, of Neotame in the
Test solution; Ai is the sum of the responses of all impurity peaks (except that of related compound A and the solvent peak, if observed) in the
Test solution; and
AS is the response of the neotame peak in the
Standard solution: not more than 2.0% is found.
Assay
Mobile phase
Dissolve 3.0 g of sodium 1-heptanesulfonate in 740 mL of water in a suitable 1000-mL vessel, add 3.8 mL of triethylamine, and mix. Adjust the resulting solution with phosphoric acid to a pH of 3.5, and dilute with water to 750 mL. Add 250 mL of acetonitrile, adjust with phosphoric acid to an apparent pH of 3.7, mix, filter, and degas. Make adjustments if necessary (see
System Suitability under
Chromatography 621).
Standard preparation
Dissolve an accurately weighed quantity of USP Neotame RS in Mobile phase, and dilute quantitatively, and stepwise if necessary, with Mobile phase to obtain a solution having a known concentration of about 1.0 mg per mL.
Assay preparation
Transfer about 50 mg of Neotame, accurately weighed, to a 25-mL volumetric flask, dissolve in and dilute with
Mobile phase to volume, and mix. Pipet 5 mL of the resulting solution into a 10-mL volumetric flask, dilute with
Mobile phase to volume, and mix. [
NOTEThis solution is stable for up to 32 hours when stored at a temperature of 0
to 10
].
Chromatographic system (see Chromatography 621)
The liquid chromatograph is equipped with a 210-nm detector and a 4.6-mm × 10-cm column that contains packing L1. The flow rate is about 1.5 mL per minute. The column temperature is maintained at 45
. Chromatograph the
Standard preparation, and record the peak responses as directed for
Procedure: the tailing factor is not more than 2.0; and the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
Separately inject equal volumes (about 25 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in mg, of C
20H
30N
2O
5 in the portion of Neotame taken by the formula:
50C(rU / rS)
in which
C is the concentration, in mg per mL, of USP Neotame RS in the
Standard preparation; and
rU and
rS are the peak responses obtained from the
Assay preparation and the
Standard preparation, respectively.
NF24